Product: Anesthesia Machine with APL valve device, Low flow rate trolley anesthesia workstation with patient monitor oxygen concentration monitoring
Item No. 850
Specifications
Anesthesia ventilator
-
Pneumatically driven electronically controlled system
-
Built-in ventilator, 10.4” LCD display
-
With 1 patient circuit for adult (Optional:patient circuit for children)
-
5-tube flowmeters, O2 & N2O & Air;0.1L/Min ~ 10 L/Min;
-
Two vaporizers,Enflurane/Isoflurane/Sevoflurane; (Optional: Halothane)
-
Oxygen Flush: 25~75 L/Min;
-
Respiration Mode: IPPV, SIPPV(VCV), IMV,SIMV, PEEP, Manu, Sigh
-
Tidal Volume: 0~1500 mL;I:E: 8:1~1:10
-
Respiration Frequency: 1~99bpm;
-
PEEP: 1~10 hPa;
-
Ptr: -10~10 hPa;
-
CO2 absorber capacity: 1 kg
-
Lung compliance: ≤ 30mL/kPa;
-
IMV frequency: 1~12bpm
-
Inspiratory Plateau: 0~1S;
-
O2 Concentration: 21%~100%
-
Alarm parameter: upper /lower limit ofairway pressure, tidal volume exceed limitation, no tidal volume output, apnea,AC power malfunction, battery low power
Patient Monitor
-
12.1''TFT large screen of high resolution
-
ECG: 5-lead, gain:×2.5mm/mV ×5.0mm/mV×10mm/mV ×20mm/mV, analysis of ST segment and abnormal of heart rate(anti-defibrillate, anti-HF electrotome)
-
Heart Rate: 20~250bpm
-
RESP: Impedance, range: 0~100bpm
-
SpO2: 0%~100%
-
Pulse Rate: 20~250bpm
-
NIBP: SBP, DBP, MBP, 10~270mmHg interval ofautomatic measurement 1~90min adjustable
-
TEMP: 0~50ºC
-
Alarm: Audio and visual alarm
-
Tendency record 96 hours
Loop part
Evaporator
-
High-precision double evaporators configuration, enflurane or isoflurane or Sevflurane or Haloflurane
-
Concentration adjustment range: 0.5 ~ 5%VOL
-
Evaporator flow, pressure, and automatictemperature compensation function
-
Self-locking feature
Configuration
-
Vaporizer: Two vaporizers, Enflurane/Isoflurane/Sevoflurane(Option: Halothane)
-
Anesthesia gas concentration range: 0.5~5%
-
Soda lime absorption system: resistance tohigh temperature of 176 degrees Celsius.
-
ISO certificates approved
|